THA MECHANICAL PART
The mechanical parts of this series machine consist of the following major components: the base unit, pillar parts (including Upright Post pieces, Rocker Arm pieces, Locking pieces), tube parts, etc.
1. Base unit
The instrument adopts Orientation foot brake, one of most advanced structures, which is stable and it is not easy to break. During its usages, the machine can be fixed quickly with the brake structure, preventing the machine from moving forward and backward. It is convenient to take radiography in a stable and accurate situation. Note: If some faults happen, please follow as bellow:
1) It brakes effectively during the movement, but when users release the foot brake, the instrument cannot move normally. At this moment, please move the machine forward and backward slowly and brakes will be able to return to its original position, then you can push the machine easily.
2) After release the foot brake and proceed as the step 1), if the machine still can not move, please use some soft object to beat the brake block below the base unit until it can work properly. Usually it is the reason that it works for a long time, and the base of the brakes lacks lubricant. Please move the arm to the lowest position and lock until hearing a "crack" locking sound, and then gently move up a little and stop when meet resistance, then move the machine to another side. The users can see the brake system below the base unit and add oli, to make it slide freely. After that, it will work well.
3) During the movement the machine in the collision process, the caster is damaged. Please tilt the machine as the above two methods, raise the damaged side of the wheel, and remove it. Then repair or replace it.
2. Pillar parts
This machine is applied to a combined pillar and rocker arm. It can take films in every position the tube parts stop, by using the built-in tension spring and rocker arm to realize the up and down movement of tube parts. Note: If some faults happen, please follow as bellow:
1) Tube components cannot stop at a certain location as the user need.
It is caused by the long time usage of tension spring. Please tilt the machine according to the "base unit" in Step 2, then use the wrench to adjust the extension spring bolt of the base unit, rotate clockwise to increase the tension, until it reaches a balance position.
2) Locking pieces cannot work.
Check whether the head pins of both sides on chrome-plated handles is broken or not. If yes, please adjust the tension spring according to above NO.1), counter-clockwise rotate until the tension spring effect can not afford to pull up, and remove the tube parts , then remove the locking pieces to check the cushion of both sides loose or not. Pleasure repair the cushion, then install locking pieces into the arm pieces together, then install the tube parts, at last adjust the tension spring bolt until balance.
3. Tube parts
Tube parts is composed of beam collimator, rotating handle, cover and tube pieces. If some faults happen during use, such as not lighting of beam collimator lamp etc., please repair and maintain according to the corresponding sections of the manual; if there is something wrong with the tube pieces, please contact our company.