Customization: | Available |
---|---|
Syringe: | Continuous Syringe |
Application: | Intravenous Injection for Human or Pet |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Features
1.Drug Library.
2.Separate operation between the two single channels.
3.Three working modes: rate mode, time-volume mode, dosage-weight mode.
4.Automatically calibration and compatible with all standard syringes of 10ml~50ml.
5.Freely Stackable: Users can freely stack one syringe pump onto another to provide multiple Solutions,
which have a wide range of clinical applications.
6.Large LCD screen displays working status.
7.KVO and Bolus functions.
8.Save the last infusion rate automatically when there is unpredicatable switch off.
9.Transfer the message between the Nurse-call System and the pumps.
Technical parameters of double channel syringe pump:
Rate: | 50/60ml Syringe: 0.1ml/h~999.9ml/h (0.1ml/h step) 1,000ml/h~1,500ml/h (1ml/h step) 30ml Syringe: 0.1ml/h~900.0ml/h (0.1ml/h step) 20ml Syringe: 0.1ml/h~600.0ml/h (0.1ml/h step) 10ml Syringe: 0.1ml/h~300.0ml/h (0.1ml/h step) |
Bolus Rate: | 50/60ml Syringe: 1,200ml/h 30 ml Syringe: 720ml/h 20 ml Syringe: 480ml/h 10 ml Syringe: 240ml/h |
Purge Rate: | 50/60ml Syringe: 1,500ml/h 30ml Syringe: 900ml/h 20ml Syringe: 600ml/h 10ml Syringe: 300ml/h |
Volume Limit: | 0.1ml~999.9ml |
Occlusion: | High: 800mmHg ±200mmHg (106.7kPa±26.7kPa Medium: 50/600mmHg ±100mmHg (66.7kPa±13.3kPa Low: 300mmHg ±100mmHg (40.7kPa±13.3kPa ) |
Alarms: | Injection soon finish, end of injection,occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose etc. |
Power Source: | AC 100V~240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup |
KVO rate: | 1ml/h |
Fuse: | F1AL/250/60V, 2pcs inside |
Power Consumption: | 30VA |
IP Classification: | IPX1 |
Equipment Classification: | Class II, internal power supply, Type BF |
Operating Condition: | Ambient temperature: +5ºC~+40ºC; |