Customization: | Available |
---|---|
Classification: | Biochemical Apparatus |
Type: | Urine Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Performance
User-friendly System
l 10.4''large color LCD displays main menu and all parameters & color histograms in one screen
l Brief operation instruction built-in software, easy for user to view
l Separate module design, very simple mechanical structure, easy maintenance
l Self check for hardware, alarm for error
l Reagent depletion alarm for Lyse, Cleanser and Diluent, as optional function
Technical Features
l 2 counting mode: whole blood, pre-dilution blood
l Sample pre-dilution: internal diluter, add 1ml diluent automatically by the needle, no need to add diluent manually to avoid deviation.
l Calibration: automated or manual
l QC: comprehensive QC program, forming and printing the histogram automatically
l Block removal: Real time monitor, high ignition, soak, flush and back flush
Overall Performance MSLBA66 | |
Test Speed | 200 Tests/hour |
System Function | Closed or open system,support different brand reagents |
24 hours continuous working,STAT function, automatic retesting | |
Dual wavelength testing | |
Test Range | Clinical biochemistry, Immunoturbidimetry, Therapeutic drug monitoring,etc |
Sample Type | Serum, plasma, urine, cerebrospinal fluid, etc |
Analysis Method | End-point,Rate,Fixed-time |
Calibration Method |
Linear Calibration:K factoring method, linear method (including two point linear and multipoint linearity) Non-Linear Calibration: Spline,Polygon,Index,Ogarithm,Logit-4P, Logit-5P |
Sample System | |
Sample Tray | 37 sample positions with cooling function |
Sample Volume | 2~70µl |
Sample Probe |
High-accuracy inside and outside wall polishing |
Liquid level sensor,anti - collision protection | |
Automatically wash interior and exterior; cross contamination rate <0.05% | |
Sample Tube |
Micro sample cups, original blood collection tubes, plastic test tubes, etc |
Reagent System | |
Reagent Tray | 28 reagents positions,analyzing maximally 45 items at the same time |
Reagent Volume | 20~350µl |
Reagent Probe |
High-accuracy inside and outside wall polishing |
Liquid level sensor,anti - collision protection | |
Automatically wash interior and exterior; cross contamination rate <0.05% | |
Reagent Refrigeration |
2~8 degree cooling or 24 hours. |
Reaction System | |
Reaction Tray | 48 reusable cuvettes (optical path length 6mm) |
Cuvettes | Plexiglass cup(optional),optical path 6mm |
Reaction Volume | 120~500µl |
Temperature Control |
37±0.1ºC,real-time monitor |
Mixing Unit | Independent muddle, nano technology, highly polished |
Optical System | |
Light Source | 12v/30w Halogen-tungsten lamp;2000 hours lifetime |
Spectral Mode | Post-spectral technology |
Wavelength | 340~800nm |
Wavelength Accuracy | <±1nm |
Absorbance Range | Linear range:0~3.0Abs |
Resolution | 0.0001 |
Stray Light | Abs>4.5 |
Data Management | |
Software Function | Friendly, easy useing, LIS system available |
Data Processing |
Full response curve and data display function, with real-time cup blank deduction, sample blank deduction, dirty cup detection; automatic re-examination of abnormal samples; sample dilution detection; real-time display of reaction plate status |
Printer | Support multiple report formats |
System Interface | Network port |
Working Conditions | |
Power Supply | AC110/220V±10%, 60/50Hz |
Power | <300VA ( UPS1000VA Optional use ) |
Ambient Temperature |
10ºC~30ºC |
Water Consumption |
Deionized water<5L/H |
Size | 950mm×610mm×510mm |
Weight | 75kg |
Computer Configuration |
|
Operating System | Windows 7/8/10, 32-bit or 64-bit version |
CPU | CPU≥2.8GHz |
RAM | RAM≥2G |
Hard Disk | Hard disk≥80 |
Monitor | 1280×102417 inches or more, resolution no less than 1280×1024 |
Interface | TCP/IP |
Printer |
Inkjet or laser printer |