Customization: | Available |
---|---|
Type: | Electric Knife |
Application: | Microsurgery |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Key Features
Maximum output power: 300W
Six output modes: (1) pure cut, (2) mixed cut 1, (3) mixed cut 2, (4) electrocoagulation 1, (5) electrocoagulation 2, (6) Bipolar condensate;
Electric cut mode: pure cut 100% cut, mixed cut 1 70 % cut, 30% coagulation; mixed cut 2 50 % cut, 50% coagulation;
Pure cut: Maximum output power 300W, enabling easy surgery on any tissues and structures;
Mixed cut: Applicable for any tissue, good coagulation effect when conducting cutting;
Electrocoagulation 1: Applicable for general open surgery and fine surgical point coagulation, featuring deep tissue coagulation layer and excellent coagulation effect;
Electrocoagulation 2: Applicable for general open surgery and fine surgical surface coagulation;
Bipolar coagulation: When the tissues achieve the effect of coagulation, the output power will begin to reduce slowly to achieve good coagulation effect;
Panel operation: Overall microcomputer automatic control, three-zone (LED) digital display, key function/power, switch/ settings, waterproof panel design;
Automatic adjustment: Automatically fine adjust power in a certain range to ensure the consistency of the cutting effect;
Automatic recognition: Neutral unipolar/bipolar plate automatic recognition, when using the dual-circuit plates and the contact area less than 60%, it should automatically stop the power output and send sound-light alarm until the contact area returns to normal, which could effectively avoid burning patients by plates.
Application range:Used in endoscopic, hysteroscopy, anorectal, general surgery, etc
Safety standard | I-type BF-type | |
Ambient temperature range | +10ºC~ +40ºC | |
Relative humidity range | 30 ~ 75% | |
Atmospheric pressure | 700~ 1060 hPa | |
Power supply | 220V ± 22V 50Hz ± 1Hz | |
Operating frequency | 530KHz ±2 | Load: Monopolar cut coagulation 300Ω, Bipolar coagulation 100Ω |
Rated output power | a)Pure cut: 1W ~ 300W(Peak voltage 1500V) | b) Mixed cut 1: 1W ~ 250W(Peak voltage 2000V) |
c)Mixed cut 2: 1W ~ 200W(Peak voltage 2500V) | d) Electrocoagulation 1: 1W ~ 120W(Peak voltage 3500V) | |
e) Electrocoagulation 2: 1W ~ 100W(Peak voltage 4020V) | f) Bipolar coagulation: 1W~50W(Peak voltage 620V) | |
Overall energy consumption | ≤1400VA | |
Transport and storage | Ambient temperature range: -40ºC ~ +55ºC | b) Relative humidity range: ≤93% |
Atmospheric pressure: 500hPa ~ 1060hPa |