Digital Mammography System | |||
High voltage generator | Type | High frequency inverter type | |
X-ray tube voltage | 20-40kV(step 0.5kV) | ||
X-ray tube current | 10mA-160mA | ||
Maximum output power | 5kW(40kV,125mA) | ||
Nominal power | 4.8kW(30kV,160mA,0.1s) | ||
Radiograph time | 0.005-10s | ||
mAs | 0.5mAS~560mAs | ||
Minimum mAs | 0.5mAs(50ms 10mA) | ||
Maximun mAs | 560mAs | ||
Minimum mAs in AEC | 6.3mAs | ||
Exposure control | manual ,semi-auto,auto | ||
Radiating mode | cross-ventilation | ||
X-ray tube | XM15 | HMT-113 | |
small focus/big focus | small focus/big focus | ||
Focus size | 0.1mm/0.3mm | 0.1mm/0.3mm | |
Maximum anode heat capacity | 225kJ(300kHU) | 225kJ(300kHU) | |
Maximum heat dissipation | 500W | 500W | |
Anode cooling mode | oil cooling | oil cooling | |
Nominal voltage | 40kV | 40kV | |
Target angle | 15° | 10°/16° | |
Target material | Mo | Mo | |
Radiation field | 18cm-50cm | 18cm-50cm | |
Inherent filtration | 0.5mmBe | 0.5mmBe | |
Additional filtration | 0.03mmMo | / | |
Anode speed | low-speed3000r/min; high-speed 10000r/min | low-speed3000r/min; high-speed 10000r/min | |
Anode input power | 1.15-2KW / 4.8-9kW | 0.7-1.35kW / 2.9-4.9kW | |
X-ray tube socket (C339v) | |||
Additional filtration | 0.03mmMo | ||
Weight | 13kg | ||
C Arm frame | |||
vertical range of motion | 630-1400mm,deviation±5% | ||
Rotation range of motion | +195°~-155°Angular division value 1°,deviation±2% | ||
FID | 660mm,deviation±5% | ||
3D tomographic axial rotation | +30°~-30°deviation±2% | ||
Rotate Mode | power-drive double speed rotate | ||
Preset angle and memory function | |||
Oppression plate | range of motion | 245mm,deviation±10% | |
range of motion in 1.5 times platfo | 150mm,deviation±10% | ||
range of motion in 1.8 times platfo | 75mm,deviation±10% | ||
Operation mode | Manual/Auto | ||
Pressure range | 0-300N(Manual)150-200N(Auto) | ||
FPD | AXS-2430 | HMD-2430 | |
Pixel unit | 85um | 85um | |
Effective pixel size | 239.36*304.64mm | 239.36*304.64mm | |
Pixel matirx | 2816*3584 pixels | 2816*3584 pixels | |
Spatial resolution | ≥7 lp/mm | ≥7 lp/mm | |
Cooling mode | fan cooling | fan cooling | |
Grayscale | ≥16bit | ≥16bit | |
DQE | ≥50% | ≥50% | |
Grid | |||
Grid proportion | 5:1 | ||
Size | 338*264mm | ||
line pair | 36lp/cm | ||
focus | 66cm | ||
Collimator (HMC-915) | |||
Power supply | 24V DC,2A | ||
Radiation field | Min 5*5cm/Max 24*30cm | ||
Radiograph platform | Field 24*30cm SID 660mm,deviation±5% | ||
1.5 times radiograph platform | Field 16*20cm SID 440mm,deviation±5% | ||
1.8 times radiograph platform | Field 13.6*16.7cm SID 366.7mm,deviation±5% |
Inherent filtration | 0.02mmAl | |
Additional filtration | 0.03mmMo /0.025Rh | |
Illumination in max SID | >160Lux | |
Timer | Button/Auto,30s timer | |
Indicator light | LED | |
maximum radiation leakage | <40mRh | |
Light field switch | Auto | |
Weight | 3.6kg | |
Image Acquisition Workstat | Screen size/resolution | 19",Resolution 1280*1024,Contrat 1000:1 |
Memory | ≥8G | |
Hard disk | ≥320G | |
OS | Win7 | |
Image format | DICOM3.0(can connect with RIS) | |
Image storage capacity | ≥20000 images | |
mamual input patients' date or download form HIS/RIS through DICOM Worklist | ||
CD-R writer (storage image) | ||
Choosing Breast implants Function | ||
Integrated design with High voltage generator, exposure specification setting in workstation | ||