Customization: | Available |
---|---|
Certification: | CE, ISO13485 |
Transport Package: | Standard Export Packing |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Fully support perspective, gastrointestinal spot film, GI (barium meal, barium enema), orthopedic photography, pediatrics photography, chest, urinary system angiography, peripheral angiography operations, gynecological Photography (HSG) and many other checks, MSLPLD6500 can realize all these functions.
Item | Content | Technical Parameters |
Power Supply | Voltage | 380V±38V |
Frequency | 50Hz±1Hz | |
Capacity | ≥65kVA | |
Internal Resistance | ≤0.17Ω | |
Generator | Power | 80KW |
Inverter Frequency | 440 KHz | |
Radiography Tube voltage | 40kV~150kV step adjustment | |
Radiography Tube current | 10mA~1000mA step adjustment | |
Radiography Time | 1.0ms~10000ms step adjustment | |
Radiography mAs | 0.1 mAs~1000 mAs | |
Fluoroscopy Tube voltage | 40kV~125kV step adjustment | |
Fluoroscopy Tube current | 0.5mA~10mA continuous fluoroscopy | |
5mA~20mA Pulse fluoroscopy | ||
X-ray tube assembly | Target | Molybdenum-based lanthanum-tungsten composite |
Target angle | 12° | |
Nominal tube voltage | 150kV | |
Tube Focus: big/small | 1.2mm / 0.6mm | |
Input power | Big Focus/ 100kW Small focus/ 40kW | |
Anode thermal capacity | 420kJ(600KHU) | |
Anode maximum heat dissipation | 1750W(2465KHU/min) | |
Component heat capacity | 1420kJ(2000KHU) | |
Anode Rotation Speed | 9700rpm | |
Collimator | Collimator field of view light | Halogen lamp, AC24V/100W |
Visible light illumination | Average illumination brightness: >100Lux | |
Light field exposure time | 5-45s, 5s per step | |
Diagnostic table |
Table transverse movement distance | 300mm |
SID | 1100mm~1800mm | |
Table panel equivalent filtration | ≤1.2mmAl | |
Point device movement range | 1250mm | |
Table height from the ground | 905mm | |
Table size | 2100mm×730mm | |
Table rotation | +90°~0°~ -25° | |
X-ray tube column swing | +40°~0°~ -40° | |
Minimum distance of the compressor from the table surface | ≤150mm | |
Compressor pressure | 80N~130N | |
Table bearing | 135kg | |
Fragmentation | Whole film, two-piece, four-slice | |
Fixed grid | Grid density: 103L/INCH, Ratio: 10:1, convergence distance 130 cm stationary type :15" ×18" |
|
Intensifier | Field of view size | 230mm |
Resolution | 52 Lp/cm | |
Conversion factor | 26 | |
X-ray camera | Signal to noise ratio | 45 dB |
Cell size | 5.86×5.86μm | |
Number of valid pixels | 1920×1200 | |
Flat Panel Detector | Image Sensor | Amorphous silicon thin film transistor |
Pixel size | 143μm | |
Effective pixel size | 3008(H)*3072(V) | |
Effective area(H x V) | 430(H)*439(V) | |
Gray scale | 14bit | |
Spatial resolution | 3.7 Lp/mm(CsI) | |
Energy range | 40kV -150kVP | |
Power input | DC24V | |
Image acquisition workstation |
Computer system | 1, computer: B150 PLUS motherboard; processor: Intel Core I5 6400 Memory: 8 GB DDR3 Hard disk: 1T SATA (7200 rpm) Optical drive: DVR-221CHV CD burner NIC: Intel(R)gigabit CT Graphics card: gtx1050 |
Monitor | Dell (DELL) U2412Mc display | |
Workstation software | Basic functions: new patient; new check; anonymous registration; WorkList search and advanced search; WorkList refresh; WorkList information import; check list; view image; film print; The collection part is divided into gastrointestinal collection and DR collection functions: Gastrointestinal collection function: bed selection; patient shape selection; specific examination protocol selection; gastrointestinal and contrast protocol selection; selected protocol list; real-time image noise reduction selection; real-time image enhancement selection; single frame storage; continuous perspective storage; Frame exposure storage; sequence exposure storage; real-time image horizontal flipping; real-time image vertical flipping; generator perspective state switching; perspective time accumulation; fluoroscopy time clearing; ABS function switch; DR acquisition function: window width window position adjustment; 90° rotation to the left; 90° rotation to the right; horizontal mirroring; vertical mirroring; image negative; cropping frame; exposure parameter setting; stadium focus switching; exposure mode selection; ionization chamber selection ; exposure actual dose value; system status indicator. Image viewing function: image zooming; partial zooming; original size; recovery window size; image grid display; 90° left rotation; 90° rotation to the right; horizontal mirroring; vertical mirroring; text logo; line logo; angle logo; |