Good Quality X-Ray Mobile Machine Rapiscan X-Ray Machine MSL70BY
Good Quality X-Ray Mobile Machine Rapiscan X-Ray Machine MSL70BYmobile x ray machine digital dr system for hospital
70mA mobile medical x-ray unit is designed for inconveniently-moved patients to take x-ray photography in the emergency room, wards, operating rooms, etc.
X-ray machine can be moved near to sickbed, operating table, stretcher, etc. The photograph diagnosis can also be taken at home or the outdoors. It is an ideal diagnostic equipment for both medical staff and Heavy jing patients.
X-ray machine is equipped with full-wave rectification combined type tube assembly. High-pressure silicon full-wave rectification technology is adopted in the tube assembly to improve x-ray tube's efficiency, also with function of temperature overload protection. There are 3 factors capacity interlock control system: kilovoltage (kV), milliampere (mA) and time(s). Controlled silicon zero passage circuit is adopted for high-voltage primary output, with overloading indicator, loading prepare indicator & the protection circuit of excessive voltage & insufficient voltage, thus the tube assembly won't be damaged by mis-operation. Cable manual switch exposure or wireless remote control emitter exposure can be chosed according to requirements.
There is acousto-optic indicator when radiant exposure which provide convenience for patients & medical personnel. Panel of this machine:
Power voltage V and tube current mA are displayed in pointer type, kilovoltage (kV) and time (s) are selected by settled grades.
The tube assembly on the cantilever of x-ray machine can be moved up and down and can be fixed at any position. It can also be settled dexterously at required angle and height to facilitate the photograph. If the whole x-ray machine need to be moved,please put the cantilever to the lowest position and lock it, so that the machine can be moved safely and conveniently.
This machine is equipped with the dynamoelectric beam limiting device with controlled radiation field position and setting amplitude of the various X-ray film. The dynamoelectric cluster limiter will send out a beam of visual light while operating to aim the nidus and the scale of the visual light overlaps the scale of X ray exposing. This function makes it convenient that medical personnel position the nidus & the x ray exposed area.
Weight of the whole machine is 97 kg. When be moved, the outside measurement is 125cm×71cm×128cm (length×width×height). It can be used to move in the ward and between beds conveniently.
The maximum output of x-ray machine is 70kV 70mA 1.0s or 90kV 30mA 5.0s; input power with single phase AV 180V~250V; maximum power consumption of 5.5kW(instantaneous); maximum photograph width(when focus-skin distance is 1m)to 40cm×40cm.
Main Technical Parameters:
1. X-ray tube
1.1 model:XD3-3.5/100
1.2 material:tungsten
1.3 reference axis of target angle and focus: target angle and focus to anode head face 95±1.75mm
1.4 angle of target surface to regulated reference axis:19°
1.5 nominal value of reference axis:2.6
1.6 inherent filtration: 0.8mmAl
1.7 nominal voltage: 100kV
1.8 heater supply: AC 6~8V Current 4.5A Frequency 50Hz
2. Tube assembly
2.1 full-wave rectification, oil cooling, combined & occluded
2.2 nominal electric power : 90kV 30mA 5.0s;70kV 70mA 1.0s
2.3 measurement size: 370mm×300mm×250mm
2.4 weight: 27kg
2.5 loading factor value of leakage radiation: 90kV 50mA 0.2s <0.02mGy/h
2.6 tube assembly belongs to I class B type, non-watertight equipment
2.7 inner surface temperature should not exceed 55ºC
2.8 filtration: 2.8mmAl
2.9 power supply of high voltage: AV 116V~214V
2.10 heater supply: AC voltage<220V
3. Electric beam limiting device
3.1 Power supply: AC 24V Capacity: 110W
3.2 light source of indicating light field: AC 24V Power 100W, tungsten halogen lamp
3.3 light field illumination intensity:>100lx when tube assembly focus to image is 100cm
3.4 width of light field:
a) when distance from x-ray tube focus to image receipt face (SID)is 65cm,max light field /axis field≤35cm×35cm
b) when distance from x-ray tube focus to image receipt face (SID)is 100cm,min light field /axis field≤5cm×5cm
3.5 filtration: 0.5mmAl
3.6 weight: 2.2kg
4.Condition of the power supply
4.1 Single-phase three-wire alternating current voltage: 180V-250V frequency: 50Hz
power supply resistance ≤0.8Ω capacity: 5.5kW instantaneous electric current: 24A (Max)
If not in this range, please select another AC voltage regulator with capacity more than 5kW.
4.2 Single phrase three-core three-core no less than 250V/16A; earth lead in the socket should be good
and reliable;qualified breaker should be placed in the front of the socket.
4.3 To promise the machine can work normally, this machine power is not possible with other large
capacity electric equipment share a power supply socket.
5. Operating condition of X-ray machine
5.1 ambient temperature:+10ºC ~+40ºC ;
5.2 relative humidity:30%~75%
5.3 atmospheric pressure:700hpa~1060hpa;
5.4. power supply:single phrase AC, voltage:220V±22V;Frequency:50Hz±1Hz;
Power:>5.5kVA;Current:>24A (instantaneous)
6. X-ray machine
6.1 Input power: 5.5kVA
6.2 Selection of timing range:0.063s~5.0s 20 digital type (execute R′10 value of GB9706.3 affix BB)
6.3 maximum rated capacity (see diagram 1)
6.4 Specification of fuse plug:6G or 6C 125V/250V φ6×32mm
6.5 Top position from focus to ground (window down): 172cm
lowest position from focus to ground (window down): 50cm
6.6 Height of safety self-lock device when x-ray machine moves up and down(from focus to
ground,window down): I grade 155cm;II grade 117cm;III grade 74cm
6.7 Width of tube assembly rotate around the axis of cantilever>90°;
width of tube assembly rotate forth around the axis of supporting stand:>90°
6.8 Mode of operation:interval loading, continuous operation
6.9 Exposure distance of remote control switch: 6m
6.10 Size of the whole machine(length ×width ×height)1510mm×710mm×1920mm
6.11 Weight of the whole machine: 97kg
6.12 Required pulling force to move the whole machine: ≤250N
6.13 Drain current, earth resistance, dielectric strength:qualified with requirements of GB9706.1,
GB9706.11